Key Benefits:
On 26 August 2015 (State 1 Er October 2015)
The Departure E Federal Justice and Police (DFJP),
Having regard to art. 5, para. 2, 8, para. 2, 11, para. 2, 16, para. 2, 17, para. 2, 24, para. 3, and 33 of the Order of 15 February 2006 on measuring instruments (OIMes) 1 ,
Stops:
This Order fixes:
1 The following are subject to the provisions of this order:
2 Electricity meters used by transient customers shall not be subject to this order:
For the purposes of this order:
Meters shall meet the essential requirements set out in Annex 1 OIMes. The following essential requirements must also be met:
Meter compliance with the essential requirements referred to in s. 4 may be evaluated and certified at the choice of the manufacturer according to one of the following procedures set out in Annex 2 OIMes:
1 Meters shall be subject to subsequent verification in accordance with Annex 7, c. 1, OIMes, carried out by the Federal Institute of Metrology (METAS) or by an authorised verification laboratory, as follows:
2 METAS may in particular cases shorten the time limits set out in para. 1 where there is a suspicion that the stability of measurement is no longer ensured before the deadline. It may order additional examinations.
3 Meters may, at the request of the user, be subject to the statistical control procedure according to Annex 4. Meters subject to the statistical control procedure shall retain their validity of the verification as long as the meters in the sample meet the requirements set out in Annex 4, let. F. Excluded from the statistical control procedure are meters which have already undergone the statistical monitoring procedure and then transferred to the control procedure according to para. 1.
Transformers must meet the essential requirements set out in Schedule 1 OIMes and Schedule 3 of this Order.
Transformers are subject to regular approval and initial verification as set out in Annex 5 OIMes.
1 Transformers are required to undergo further verification in accordance with Annex 7, c. 1, OIMes carried out by METAS or by an authorized verification laboratory, as follows:
2 To the extent permitted, METAS may set a different interval for the subsequent verification of transformers according to para. 1, let. B.
The user assumes the responsibility specified in s. 21, para. 1, OIMes, but also:
1 Measurement groups comprising meters and transformers shall be adjusted so that the energy required for the operation of the measuring device is not measured.
2 If a measurement group is to be connected in derogation from para. 1, its own consumption alone should not lead to a display of energy consumption.
3 The additional error caused by the connection cables and the loads of transformers in measurement groups shall not exceed 20 % of the maximum permissible error of the meter.
1 Users maintain a control log of the measurement instruments used in their field of activity.
2 The register shall specify for each measuring instrument:
3 The energy consumers concerned and the bodies responsible for the implementation of the subsequent control may consult the register at any time.
4 Records must be retained after the expiry of the last audit as long as there are open claims, but at least for five calendar years.
The DFJP Order of 19 March 2006 on Energy and Power Measurement Instruments 1 Is repealed.
1 [ RO 2006 1613 , 2012 7183]
1 Verification of electricity meters carried out prior to the coming into force of this order shall remain valid for the period defined by the law in force at the time of the verification.
2 Verification of measurement transformers made before the entry into force of this order shall remain valid until 31 December 2075.
3 Electricity meters and measuring transformers that have been approved prior to October 30, 2006, may be placed on the market and subject to the initial verification set out in Schedule 5, c. 2, OIMes until October 29, 2016. After the placing on the market, the maintenance of the stability of measurement shall be governed by the provisions of this order; paras. 1 and 2 remain reserved.
4 Electricity meters and measuring transformers that have been approved between October 30, 2006 and the coming into force of this order may still be placed on the market and subject to the initial verification set out in Schedule 5, Ch. 2, OIMes until approval expires. After the placing on the market, the maintenance of the stability of measurement shall be governed by the provisions of this order; paras. 1 and 2 remain reserved.
5 Electricity meters equipped with the measurement functionality of the load curve power or functionality may be placed on the market until 31 December 2017, even if they do not comply with the requirements Of this order concerning these functionalities. From 1 Er January 2018, the maintenance of the stability of measurement of these functionalities is governed by the provisions of this order. The performance of the maintenance of measure stability procedure is performed for these functionalities for the first time at the same time as the next procedure for maintaining the stability of the measure applicable to other intended functionalities From 1 Er January 2018.
6 The classes for the accuracy of active energy meters under the old right correspond to the following classes according to the new right:
7 For the measurement of reactive energy, Class 1 under the old right corresponds to Class 2 under the new right.
8 For active energy meters subject to the statistical control procedure already before the entry into force of this order:
9 The first announcement of electricity meters for the statistical monitoring procedure is possible until 31 October 2015 if the majority of electricity meters announced were manufactured in 2011. L' al. 8 shall apply mutatis mutandis.
This order shall enter into force on 1 Er October 2015.
26 August 2015 |
Federal Department of Justice and Police: Simonetta Sommaruga |
(art. 4)
I |
= |
The measurement connected to the entry of the meter, proportional to the current intensity crossing the current circuit; |
I St |
= |
Value of I Declared the lowest to which the meter registers the active electrical energy with cos ö = 1 (Balanced load polyphase meters); |
I Min |
= |
Value of I , beyond which the error is in the maximum tolerated errors corresponding to the meter accuracy class (polyphase load balanced meters); |
I Tr |
= |
Value of I Beyond which the error is in the lowest tolerated maximum errors corresponding to the meter accuracy class; |
I N |
= |
Specified reference current for which the meter supplied by a transformer was designed; |
I B |
= |
Reference current of I For which the direct connection meter was designed ( I B = 10 I Tr ); |
I Max |
= |
Maximum value of I For which the error is in the maximum tolerated errors; |
U |
= |
The measurement connected to the entries of the meter, proportional to the electrical tension in the current circuit; |
U N |
= |
Reference voltage of U ; |
F |
= |
Frequency of the voltage provided to the meter; |
F N |
= |
Reference value of F ; |
ö |
= |
Phase difference ö , between I And U . |
Class A |
Class B |
Class C |
||
Direct plugging meters |
||||
I St |
£ 0.05 I Tr |
£ 0.04 I Tr |
£ 0.04 I Tr |
|
I Min |
£ 0.5 I Tr |
£ 0.5 I Tr |
£ 0.3 I Tr |
|
I Max |
³ 50 I Tr |
³ 50 I Tr |
³ 50 I Tr |
|
Meters on Measuring Transformer |
||||
I St |
£ 0.06 I Tr |
£ 0.04 I Tr |
£ 0.02 I Tr |
|
I Min |
£ 0.4 I Tr |
£ 0.2 I Tr * |
£ 0.2 I Tr |
|
I N |
= 20 I Tr |
= 20 I Tr |
= 20 I Tr |
|
I Max |
³ 1.2 I N |
³ 1.2 I N |
³ 1.2 I N |
|
* Condition I Min £ 0.4 I Tr Applies to Class B electro-mechanical meters. |
Table 2
Operating temperatures |
|||||||||||||
+ 5 ° C to +30 ° C |
-10 ° C to + 5 ° C or +30 ° C to +40 ° C |
-25 ° C to -10 ° C or +40 ° C to +55 ° C |
-40 ° C to -25 ° C or +55 ° C to +70 ° C |
||||||||||
Meter class |
A |
B |
C |
A |
B |
C |
A |
B |
C |
A |
B |
C |
|
Single phase meter; polyphase meter if operating at balanced loads |
|||||||||||||
I Min ≤ I < I Tr |
3.5 |
2 |
1 |
5 |
2.5 |
1.3 |
7 |
3.5 |
1.7 |
9 |
4 |
2 |
|
I Tr ≤ I < I Max |
3.5 |
2 |
0.7 |
4.5 |
2.5 |
1 |
7 |
3.5 |
1.3 |
9 |
4 |
1.5 |
|
Polyphase meter if it operates on a single phase load |
|||||||||||||
I Tr ≤ I < I Max * |
4 |
2.5 |
1 |
5 |
3 |
1.3 |
7 |
4 |
1.7 |
9 |
4.5 |
2 |
|
* Exception: for polyphase electromechanical meters, the range of current intensity for single phase load is limited to 5 I Tr £ I £ I Max . |
Table 3
Disturbances |
Limit Values in % for Class Meters |
|||
A |
B |
C |
||
Reverse Phase Sequence |
1.5 |
1.5 |
0.3 |
|
Voltage disequilibrium (applicable only to polyphase meters) |
4 |
2 |
1 |
|
Harmonics in current circuits * |
1 |
0.8 |
0.5 |
|
Continuous and harmonic current in current circuits * |
6 |
3 |
1.5 |
|
Fast Transient Salves |
6 |
4 |
2 |
|
Magnetic fields, HF electromagnetic fields, conduction disturbances introduced by fields at radio frequencies and immunity to oscillatory waves |
3 |
2 |
1 |
|
* for electromechanical electricity meters, no limit value is defined for harmonics in current circuits. |
(art. 4)
Table 1
Class 3 |
Class 2 |
||
Direct Connection Counter |
|||
I St |
≤0.01 I B |
≤0.005 I B |
|
Metering Transformer Meter |
|||
I St |
≤0.005 I N |
≤0.003 I N |
|
Table 2
I |
Sin ö (inductive or capacitive) |
Limit Values in % for Class Meters |
||
Direct Connection Counter |
Metering Transformer Meter |
3 |
2 |
|
Single phase meter; polyphase balanced charge meter |
||||
0.05 I B ≤ I < 0.1 I B |
0.02 I N ≤ I < 0.05 I N |
Sin ö = 1 |
4.0 |
2.5 |
0.1 I B ≤ I ≤ I Max |
0.05 I N ≤ I ≤ I Max |
Sin ö = 1 |
3.0 |
2.0 |
0.1 I B ≤ I < 0.2 I B |
0.05 I N ≤ I < 0.1 I N |
0.5 ≤sin ö < 1 |
4.0 |
2.5 |
0.2 I B ≤ I ≤ I Max |
0.1 I N ≤ I ≤ I Max |
0.5 ≤sin ö < 1 |
3.0 |
2.0 |
0.2 I B ≤ I ≤ I Max |
0.1 I N ≤ I ≤ I Max |
0.25 ≤sin ö ≤0.5 |
4.0 |
2.5 |
Polyphase meter with single phase load |
||||
0.1 I B ≤ I ≤ I Max |
0.05 I N ≤ I ≤ I Max |
Sin ö = 1 |
4.0 |
3.0 |
0.2 I B ≤ I ≤ I Max |
0.1 I N ≤ I ≤ I Max |
0.5 ≤sin ö < 1 |
4.0 |
3.0 |
(art. 8)
Table 1
Current Range |
EMT for Transformation Report K % |
EMT for phase shift ö In minutes |
||
Class |
Class |
|||
0.5 S |
0.2 S |
0.5 S |
0.2 S |
|
I Min ≤ I P < 5 % I N |
1.5 |
0.75 |
90 |
30 |
5 % I N ≤ I P < 20 % I N |
0.75 |
0.35 |
45 |
15 |
20 % I N ≤ I P ≤ I Max |
0.5 |
0.2 |
30 |
10 |
Table 2
Current Range |
EMT for Transformation Report K % |
EMT for phase shift ö In minutes |
||||
Class |
Class |
|||||
0.5 |
0.2 |
0.1 |
0.5 |
0.2 |
0.1 |
|
I Min ≤ I P < 20 % I N |
1.5 |
0.75 |
0.4 |
90 |
30 |
15 |
20 % I N ≤ I P < 100 % I N |
0.75 |
0.35 |
0.2 |
45 |
15 |
8 |
100 % I N ≤ I P ≤ I Max |
0.5 |
0.2 |
0.1 |
30 |
10 |
5 |
Table 3
Voltage range |
EMT for Transformation Report K % |
EMT for phase shift ö In minutes |
||||
Class |
Class |
|||||
0.5 |
0.2 |
0.1 |
0.5 |
0.2 |
0.1 |
|
U Min ≤ U P ≤ U Max |
0.5 |
0.2 |
0.1 |
20 |
10 |
5 |
(art. 6, para. 3)