Advanced Search

RS 941.231 DFJP Order of 19 March 2006 on Thermal Energy Measuring Instruments

Original Language Title: RS 941.231 Ordonnance du DFJP du 19 mars 2006 sur les instruments de mesure de l’énergie thermique

Subscribe to a Global-Regulation Premium Membership Today!

Key Benefits:

Subscribe Now for only USD$40 per month.

941.231

DFJP Ordinance on Thermal Energy Measuring Instruments

19 March 2006 (State 1 Er January 2013)

Federal Department of Justice and Police (DFJP),

Having regard to art. 5, para. 2, 8, para. 2, 11, para. 2, 16, para. 2, 17, para. 2, 24, para. 3, and 33 of the order of 15 February 2006 on measuring instruments (Ordinance on measuring instruments) 1 , 2

Stops:

Section 1 General provisions

Art. 1 Purpose

This Order fixes:

A.
Requirements for hot water meters, thermal energy meters and cold meters;
B.
The procedures for placing these measuring instruments on the market;
C.
Procedures to maintain the stability of measurement of these measuring instruments.
Art. 2 Scope of application

The following are subject to the provisions of this order:

A.
Hot water meters to measure volumes of hot water used in households, the arts and crafts, or light industry;
B.
Thermal energy meters and cold meters which are intended to determine energy costs and are used in households, the arts and crafts or light industry.
Art. 3 Definitions

For the purposes of this order:

A.
Hot water meter: A measuring instrument that determines the volume of running water or sanitary water in excess of 30 ° C;
B.
Thermal energy meter: An instrument for measuring the thermal energy released in a heat exchange circuit by a coolant (water or superheated steam);
C.
Cold meter: An instrument for measuring thermal energy subtracted from water in a water-cooling system.

Section 2 Hot water meters

Art. 4 Essential requirements

Hot water meters shall meet the essential requirements set out in Schedule 1 of the Order on Measurement Devices and Schedule 1 of this Order.

Art. 5 Marketing Procedures

The compliance of hot water meters with the essential requirements referred to in s. 4 may be evaluated and certified at the manufacturer's choice in accordance with one of the following procedures set out in Schedule 2 of the Ordinance on Measuring Instruments:

A.
Type examination (module B) completed by the declaration of conformity to the type on the basis of quality assurance of the manufacturing process (module D);
B.
Type examination (module B) completed by the declaration of conformity to the type on the basis of a product verification (module F);
C.
Declaration of conformity on the basis of comprehensive quality assurance and design control (module H1).
Art. 6 Procedures for maintaining measurement stability

1 For hot water meters, the user may choose one of the following two procedures to ensure the maintenance of measurement stability:

A.
Subsequent verification in accordance with Annex 7, c. 1, of the order on measuring instruments, carried out every five years by a qualified audit laboratory;
B.
Monitoring of measuring instruments in service as per Annex 7, c. 3, of the order on measuring instruments and in accordance with the requirements set out in Annex 2, c. 1, of this Order.

2 Hot water meters used to determine the proportional distribution of energy costs are not subject to any procedures for maintaining measurement stability.

Section 3 Thermal energy meters

Art. 7 Essential requirements

1 Thermal energy meters for liquids shall meet the essential requirements set out in Schedule 1 of the Order on Measurement Devices and Schedule 3 of this Order.

2 The superheated steam thermal energy meters shall meet the essential requirements set out in Schedule 1 of the Instrument for Measurement and Schedule 4 of this Order.

Art. 8 Marketing Procedures

1 The conformity of thermal energy meters for liquids with the essential requirements referred to in Art. 7, para. 1, may be evaluated and certified at the selection of the manufacturer in accordance with one of the following procedures set out in Schedule 2 of the Ordinance on Measuring Instruments:

A.
Type examination (module B) completed by the declaration of conformity to the type on the basis of quality assurance of the manufacturing process (module D);
B.
Type examination (module B) completed by the declaration of conformity to the type on the basis of a product verification (module F);
C.
Declaration of conformity on the basis of comprehensive quality assurance and design control (module H1).

2 Thermal energy meters for superheated steam are subject to individual approval and an initial verification according to Annex 5 of the Ordinance on Measuring Instruments.

Art. Procedures for maintaining measurement stability

1 For thermal energy meters for liquids, the user may choose from the following procedures to ensure the maintenance of the stability of measurement:

A.
Subsequent verification in accordance with Annex 7, c. 1, of the order on measuring instruments, carried out every five years by a qualified audit laboratory;
B.
Monitoring of measuring instruments in service as per Annex 7, c. 3, of the order on measuring instruments and in accordance with the requirements set out in Annex 2, c. 1, of this order;
C.
Calibration according to Annex 7, c. 6, of the order on measuring instruments and in accordance with the requirements set out in Annex 2, c. 2, of this Order.

2 For over-heated thermal energy meters, the user may choose one of the following two procedures to ensure the maintenance of measurement stability:

A.
Subsequent verification in accordance with Annex 7, c. 1, of the order on measuring instruments, carried out every five years by a qualified audit laboratory;
B.
Calibration according to Annex 7, c. 6, of the order on measuring instruments and in accordance with the requirements set out in Annex 2, c. 2, of this Order.

3 Thermal energy meters used to determine the proportional distribution of energy costs are not subject to any procedure for maintaining measurement stability.

Section 4 Cold Meters

Art. 10 Essential requirements

Cold meters shall meet the essential requirements set out in 1 of the Order on Measurement Devices and Schedule 5 of this Order.

Art. 11 Marketing Procedures

Cold meters are subject to regular approval and initial verification according to Annex 5 of the Ordinance on Measuring Instruments.

Art. 12 Procedures for maintaining measurement stability

Cold meters shall be subjected to a subsequent verification every five years in accordance with Annex 7, c. 1, of the order on measuring instruments, carried out by an authorized verification laboratory.

Section 5 User obligations

Art. 13 Mounting, commissioning and maintenance of the measuring instrument

The user assumes the responsibility specified in s. 21, para. 1, of the order on measuring instruments, but also the order:

A.
Ensure compliance with the manufacturer's instructions for the assembly and commissioning of the measuring instrument;
B.
Maintain the measuring instrument in good condition and periodically review the parts subject to wear, aging or fouling.
Art. 14 Control records

1 The user maintains a control register of the measuring instruments used in his or her field of activity.

2 The register shall specify for each measuring instrument:

A.
When and according to what procedure it has been placed on the market;
B.
The procedure for maintaining measurement stability is prescribed;
C.
When the maintenance of measurement stability procedure was last applied;
D.
Where it is in service.

3 The energy consumers concerned and the bodies responsible for the enforcement of this order may consult the register at any time.

4 In case of disagreement, the Federal Institute of Metrology (METAS) 1 Decides whether the records meet the requirements.


1 The designation of the administrative unit has been adapted to 1 Er Jan 2013 under Art. 16 al. 3 of the O of 17 Nov 2004 on Official Publications (RS 170.512.1 ). This mod has been taken into account. Throughout the text.

Section 6 Maximum Tolerable Errors During Controls

Art. 15

The maximum permissible errors applicable in the challenge of results of measures within the meaning of Art. 29, para. 1, of the order on measuring instruments or in the official control of an instrument of measurement carried out outside the verification shall be the double of the maximum permissible errors for complete instruments fixed in Annexes 1, 3, 4 and 5 of the This order.

Section 7 Final provisions

Art. 16 Repeal of the law in force

The Order of 21 May 1986 on Thermal Energy Meters 1 Is repealed.


1 [RO 1986 1121, 1997 2761 hp. II let.]

Art. 17 Transitional provisions

1 Hot water meters and thermal energy meters for liquids verified prior to the coming into force of this order may continue to be verified. At the time of the subsequent verification, meters shall comply with the maximum permissible errors in accordance with the previous provisions.

2 Hot water meters and thermal energy meters for liquids approved under the earlier right may be put into service and verified initially according to Annex 5, c. 2, of the order on instruments of measurement for ten years from the coming into force of this order. After the expiry of these ten years, they may still be verified at a later date.

3 Thermal energy meters for overheated steam and cold meters placed on the market prior to the coming into force of this order may be used without verification for five years after they are commissioned or revised.

4 Meters mentioned in para. 3 may be verified at a later date if they meet the requirements of this order.

5 Thermal energy meters and hot water meters used to determine the proportional distribution of energy costs may be placed on the market in accordance with the earlier provisions for five years from the entry In force of this Order.

Art. 18 Entry into force

This Order comes into force on October 30, 2006.

Annex 1

(art. 4)

Specific requirements for hot water meters

A Definitions

Minimum Throughset (Q 1 )

The lowest flow rate for which the hot water meter displays results that meet the maximum tolerated error requirements.

Transient rate (Q 2 )

The rate between the stable flow and the minimum flow rate that is delineating two zones of the flow range, namely the upper and lower load extents to which different maximum tolerated errors apply.

Stable rate (Q 3 )

The maximum flow rate for which the hot water meter operates satisfactorily under normal conditions of use, that is, in stable or variable flow conditions.

Overload rate (Q 4 )

The maximum rate for which the meter operates satisfactorily, without disturbance, for a short time.

B Metrological Requirements

1

Nominal Operating Conditions

1.1
The manufacturer shall indicate the nominal operating conditions for the flow rate of the hot water.
The values of the flow range must meet the following conditions:
-
Q 3 /Q 1 ³ 10;
-
Q 2 /Q 1 = 1.6;
-
Q 4 /Q 3 = 1.25.
Within five years after the coming into force of this Order, the Q 2 /Q 1 May have values of 1.5; 2.5; 4 or 6.3.
1.2
The manufacturer shall indicate the nominal operating conditions for the temperature range of the hot water.
The temperature range should range from 30 ° C to at least 90 ° C.
1.3
The manufacturer shall indicate the nominal operating conditions for the relative pressure range of the hot water.
The relative pressure range should range from 0.3 bar to at least 10 bars for Q 3 .
1.4
The manufacturer shall indicate the nominal operating conditions for the power supply: nominal values of the ac power supply and/or limit values of the dc power supply.

2

Maximum Tolerable Errors

2.1
Maximum tolerated, positive or negative errors are 3 % on the volumes provided for flow rates between the transitory flow rate (Q 2 ) (included) and overload rate (Q 4 ).
2.2
Maximum tolerated, positive or negative errors are 5 % on the volumes provided for flows between the minimum flow rate (Q 1 ) And the transient rate (Q 2 ) (excluded).

3

Electromagnetic immunity

3.1
An electromagnetic disturbance cannot influence the hot water meter:
-
If the change in the result of the measure does not exceed the limit value specified in the c. 3.3, or
-
If the indication of the result of the measure is such that it cannot be regarded as valid, as is the case in a short fluctuation which cannot be interpreted, recorded or transmitted as a result of the measure.
3.2
After an electromagnetic disturbance, the hot water meter shall:
-
Resume operation within the limits of maximum tolerated errors;
-
Have all of its measurement functions insured, and
-
To recover all available measurement data just before the disturbance.
3.3
The limit value is the smallest of the following two values:
-
The volume that corresponds to one-half of the maximum permissible error on the volume measured for the upper load range;
-
The volume that corresponds to the maximum tolerated error for the volume that transited in one minute at flow Q 3 .

4

Sustainability

4.1
After an appropriate review has been carried out taking into account the time limits proposed by the manufacturer, the following criteria must be met:
4.2
At the end of the durability test, the variation of the measurement against the result of the initial measurement shall not exceed the following values:
-
± 3 % of the volume measured between Q 1 (included) and Q 2 (excluded);
-
± 1.5 % of the volume measured between Q 2 (included) and Q 4 (included).
4.3
The measurement deviation for the volume measured after the durability test shall not exceed the following values:
-
± 6 % of the volume measured between Q 1 (included) and Q 2 (excluded);
-
± 3.5 % of the volume measured between Q 2 (included) and Q 4 (included).

5

Adéquation

5.1
Unless otherwise indicated, the meter shall be capable of being used in all positions.
5.2
The manufacturer must specify whether the meter is designed to measure inverted flows. If this is the case, the reverse flow volume must be either subtracted from the cumulative volume or entered separately. The maximum tolerated error is the same for the normal flow and the reverse flow.
5.3
Hot water meters which are not designed to measure reverse flow shall either prevent or withstand an accidental spill without damage or modification of their metrological characteristics.

6

Units of Measure

The measured volume should be displayed in m 3 .

State 1 Er January 2013

Annex 2

(art. 6 and 9)

Procedures for maintaining measurement stability for hot water meters and thermal energy meters

1

Monitoring of measuring instruments in service

If the user makes use of at least 150 hot water meters and thermal energy meters, the user may apply to METAS for a procedure to extend the verification period under the following conditions:
-
The procedure should ensure proper measurements of correct measurements;
-
All hot water meters and thermal energy meters in service must have been placed on the market in accordance with s. 5 or 8 of this order, and none of these meters, at any time, may have been in service for more than ten years without further verification under Schedule 7, c. 1, of the Ordinance on Measuring Instruments;
-
Defective meters shall be replaced by conforming meters;
-
All meters in service must be operated under comparable conditions;
-
The user informs METAS once a year of the results of the procedure.

2

Calibration

If the rated power of the thermal energy metering installation is at least 10 MW, the user may choose the calibration according to Annex 7, c. 6, of the order on measuring instruments to ensure the maintenance of stability of measurement if the following conditions are met:
-
The exchange of thermal energy takes place by fixed pipes between two permanent partners, the quantity supplied being determined in a measurement station on the basis of the data supplied by one or more thermal energy meters, including the Sum of the nominal powers is at least 10 MW;
-
The thermal energy metering installation is in place between two trading partners, both of whom are able to judge the outcome of the measures;
-
The thermal energy metering installation is subject to regular metrological monitoring of qualified operating personnel;
-
If parts of the metering installation cannot be calibrated in the undertaking, the calibration of the measuring instruments shall be performed by a calibration laboratory recognized by METAS, by an authorized verification laboratory or by METAS. The installation of thermal energy measurement shall be sealed in an identifiable manner after calibration;
-
The thermal energy metering installation is calibrated as required, but generally every 12 months. The validity of the calibration of each part of the metering installation shall not exceed two years;
-
All work (maintenance, adjustment, calibration) performed on the thermal energy metering installation is recorded in minutes. The data collected must be used to determine what work has been done, when and by whom. In the event of a dispute, the minutes must be able to be submitted to the appropriate body.

State 1 Er January 2013

Annex 3

(art. 7, para. 1)

Specific requirements for thermal energy meters for liquids

A Definitions

A thermal energy meter for liquids is either a complete meter or a combined meter consisting of the subsets "flow sensor", "pair of temperature sensors" and "unit of calculation" or a combination thereof.

Q

=

Temperature of the coolant;

Q In

=

Value of Q At the inlet of the heat energy exchanger;

Q Out

=

Value of Q Return of the heat energy exchanger;

Q

=

Temperature difference Q In - Q Out With ∆ Q ³ 0;

Q Max

=

Upper limit of Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q Min

=

Lower limit of Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q Max

=

Upper limit of ∆ Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q Min

=

∆ lower limit Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q

=

The flow of the coolant;

Q S

=

Upper value of q briefly allowable when the thermal energy meter is working properly;

Q P

=

Maximum value of q allowed on a permanent basis when the thermal energy meter is working properly;

Q I

=

Minimum value of q allowed when the thermal energy meter is working correctly;

P

=

Heat power exchanged;

P S

=

Upper limit of P allowed when the thermal energy meter is functioning properly.

B Metrological Requirements

1

Nominal Operating Conditions

1.1
The manufacturer shall indicate the nominal operating conditions for the temperature of the liquid: Q Max , Q Min .
1.2
The manufacturer shall indicate the nominal operating conditions for the temperature difference of the liquid: ∆ Q Max , ∆ Q Min , the following limitations are imposed: ∆ Q Max /∆ Q Min ³ 10; ∆ Q Min = 3 K or 5 K or 10 K.
1.3
The manufacturer shall indicate the nominal operating conditions for the liquid flow: q S , q P , q I , the following limitation is imposed on the values of q P And q I : q P /q I ³ 10.
1.4
The manufacturer shall indicate the nominal operating conditions for the heat power: P S .

2

Accuracy Classes

The following accuracy classes are fixed for thermal energy meters: 1 2 3.

3

Maximum Tolerable Errors for Full Thermal Energy Meters

The maximum tolerated relative errors for the full thermal energy meters expressed as % of the true value are for each accuracy class:
-
E = E F + E T + E C .
E F , E T And E C Are set to c. 7.1 to 7.3.

4

Electromagnetic immunity

4.1
The instrument of measurement shall not be influenced by a static magnetic field or an electromagnetic field at the frequency of the network.
4.2
An electromagnetic disturbance can influence the thermal energy meter only:
-
If the change in the result of the measure does not exceed the limit value specified in the c. 4.3, or
-
If the indication of the result of the measure is such that it cannot be regarded as valid.
4.3
The limit value for complete thermal energy meters is equal to the absolute value of the maximum permissible error fixed for this thermal energy meter (see c. 3).

5

Sustainability

5.1
After an appropriate review has been carried out taking into account the time limits proposed by the manufacturer, the following criteria must be met:
5.2
Flow sensors: at the end of the durability test, the variation of the measurement relative to the result of the initial measurement shall not exceed the limit value.
5.3
Temperature sensors: at the end of the durability test, the measured values shall not differ from the initial values of more than 0.1 ° C.

6

Thermal Energy Meter Enrolment

-
Class of accuracy;
-
Flow limit values;
-
Temperature limit values;
-
Temperature difference limit values;
-
Flow sensor installation location: entry or return;
-
Indication of flow direction.

7

Subsets

Subset provisions may apply to sub-assemblies manufactured by one or more manufacturers. The essential requirements for thermal energy meters shall apply equally to sub-assemblies. The following additional requirements apply:
7.1
Relative maximum tolerable error for flow sensor expressed as % in accuracy classes:
-
Class 1: E F = (1 + 0.01 q P /q), but not more than 5 per cent;
-
Class 2: E F = (2 + 0.02 q P /q), but not more than 5 per cent;
-
Class 3: E F = (3 + 0.05 q P /q), but not more than 5 %.
E F Is the difference between the specified value and the true value of the relationship between the output signal of the flow sensor and the mass or volume.
7.2
Relative maximum permissible error for pair of temperature sensors expressed as %:
-
E T = (0.5 + 3-∆ Q Min /∆ Q ).
E T Is the difference between the indicated value and the true value of the relationship between the output signal of the temperature sensor pair and the temperature difference.
7.3
Relative maximum tolerable error for calculation unit expressed as %:
-
E C = (0.5 + ∆ Q Min /∆ Q ).
E C Is the difference between the value of the specified heat energy and the true value.
7.4
The limit value for a subset of a thermal energy meter is equal to the absolute value of the maximum permissible error for the subset (see ch. 7.1, 7.2 or 7.3).
7.5
Subset subscriptions

Debit sensor:

-

Class of accuracy;

-

Flow limit values;

-

Temperature limit values;

-

Pulse value (p. Ex. Litre/pulse) or corresponding output signal;

-

Indication of flow direction;

Temperature Sensor Pair:

-

Sensor type (p. Ex. Pt 100);

-

Temperature limit values;

-

Temperature difference limit values;

Calculation unit:

-

Type of temperature sensors;

-

Temperature limit values;

-

Temperature difference limit values;

-

Pulse value (p. Ex. Litre/pulse) or corresponding input signal emitted by the flow sensor;

-

Flow sensor mounting mode: input or return.


State 1 Er January 2013

Annex 4

(art. 7, para. 2)

Specific requirements for over-heated steam thermal energy meters

A Definitions

An overheated steam thermal energy meter is a complete meter consisting of the subassemblies "flow sensor", "pair of temperature sensors" and "unit of calculation".

Q

=

Temperature of the coolant;

Q In

=

Value of Q At the inlet of the heat energy exchanger;

Q Out

=

Value of Q Return of the heat energy exchanger;

Q

=

Temperature difference Q In - Q Out With ∆ Q ³ 0;

Q Max

=

Upper limit of Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q Min

=

Lower limit of Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q Max

=

Upper limit of ∆ Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q Min

=

∆ lower limit Q For the correct operation of the thermal energy meter within the limits of the maximum permissible errors;

Q

=

Rate of the coolant;

Q S

=

Maximum value of q briefly allowable when the thermal energy meter is functioning properly;

Q P

=

Maximum value of q allowed on a permanent basis when the thermal energy meter is working properly;

Q I

=

Minimum value of q allowed when the thermal energy meter is working correctly;

P

=

Heat power exchanged;

P S

=

Upper limit of P allowed when the thermal energy meter is functioning properly.

B Metrological Requirements

1

Nominal Operating Conditions

1.1
The manufacturer shall indicate the nominal operating conditions for the temperature of the coolant: Q Max , Q Min .
1.2
The manufacturer shall indicate the nominal operating conditions for the temperature difference of the coolant: ∆ Q Max , ∆ Q Min .
1.3
The manufacturer shall indicate the nominal operating conditions for the overheated steam flow: q S , q P , q I .
1.4
The manufacturer shall indicate the nominal operating conditions for the heat power: P S .

2

Maximum Tolerable Errors for Full Thermal Energy Meters

The maximum allowable errors for the full thermal energy meters expressed as % of the true value are as follows:
-
E = 4 + 4-∆ Q Min /∆ Q + 0.05 q P /q, but at least E = 6 + 4-∆ Q Min /∆ Q .

3

Electromagnetic immunity

3.1
The instrument of measurement shall not be influenced by a static magnetic field or an electromagnetic field at the frequency of the network.
3.2
An electromagnetic disturbance can influence the thermal energy meter only:
-
If the change in the result of the measure does not exceed the limit value specified in the c. 3.3, or
-
If the indication of the result of the measure is such that it cannot be regarded as valid.
3.3
The limit value for complete thermal energy meters is equal to the absolute value of the maximum permissible error fixed for this thermal energy meter (see c. 2).

4

Sustainability

At the end of an appropriate inspection review carried out taking into account the time limits proposed by the manufacturer, the variation of the result of the measure in relation to the result of the initial measure must not exceed the maximum tolerated errors (see ch. 2).

5

Thermal Energy Meter Enrolment

-
Flow limit values;
-
Temperature limit values;
-
Temperature difference limit values;
-
Flow sensor installation location: entry or return;
-
Indication of flow direction.

State 1 Er January 2013

Annex 5

(art. 10)

Specific requirements for cold meters

A Definitions

A cold meter is either a complete meter or a combined meter consisting of the subsets "flow sensor", "pair of temperature sensors" and "unit of calculation" or a combination thereof.

Q

=

Temperature of the coolant;

Q In

=

Value of Q At the entrance to the cooling circuit;

Q Out

=

Value of Q The return of the cooling circuit;

Q

=

Temperature difference Q In - Q Out With ∆ Q ≤0;

Q Max

=

Upper limit of Q For the correct operation of the cold meter within the limits of the maximum permissible errors;

Q Min

=

Lower limit of Q For the correct operation of the cold meter within the limits of the maximum permissible errors;

Q Max

=

Upper limit of ∆ Q For the correct operation of the cold meter within the limits of the maximum permissible errors;

Q Min

=

∆ lower limit Q For the correct operation of the cold meter within the limits of the maximum permissible errors;

Q

=

Flow of the refrigerated liquid;

Q S

=

Maximum value of q briefly allowable when the cold meter is working properly;

Q P

=

Maximum value of q allowed on a permanent basis when the cold meter is functioning properly;

Q I

=

Minimum value of q allowed when the cold meter is working correctly;

P

=

Refrigerated cooling power

P S

=

Upper limit of P allowed when the cold meter is working correctly.

B Metrological Requirements

1

Nominal Operating Conditions

1.1
The manufacturer shall indicate the nominal operating conditions for the temperature of the refrigerated liquid: Q Max , Q Min .
1.2
The manufacturer shall indicate the nominal operating conditions for the temperature difference of the refrigerated liquid: ∆ Q Max , ∆ Q Min .
1.3
The manufacturer shall indicate the nominal operating conditions for the liquid flow: q S , q P , q I, Q values P And q I Being limited as follows: q P /q I ³ 10.
1.4
The manufacturer shall indicate the nominal operating conditions for the refrigeration capacity: P S .

2

Accuracy Classes

The following accuracy classes are fixed for cold meters: 2 3.

3

Maximum toleration errors for complete cold meters

The maximum tolerated relative maximum errors for full cold meters expressed as % of the true value are for each accuracy class:
-
E = E F + E T + E C .
E F , E T And E C Are set to c. 7.1 to 7.3.

4

Electromagnetic immunity

4.1
The instrument of measurement shall not be influenced by a static magnetic field or an electromagnetic field at the frequency of the network.
4.2
An electromagnetic disturbance may influence the cold meter only:
-
If the change in the result of the measure does not exceed the limit value specified in the c. 4.3, or
-
If the indication of the result of the measure is such that it cannot be regarded as valid.
4.3
The limit value for complete cold meters is equal to the absolute value of the maximum permissible error fixed for this cold meter (see c. 3).

5

Sustainability

5.1
After an appropriate review has been carried out taking into account the time limits proposed by the manufacturer, the following criteria must be met:
5.2
Flow sensors: at the end of the durability test, the variation of the result of the measure in relation to the result of the initial measurement shall not exceed the limit value.
5.3
Temperature sensors: at the end of the durability test, the variation of the result of the measure with respect to the result of the initial measurement shall not exceed 0.1 ° C.

6

Enrolment on the cold meter

-
Class of accuracy;
-
Flow limit values;
-
Temperature limit values;
-
Temperature difference limit values;
-
Flow sensor installation location: entry or return;
-
Indication of flow direction.

7

Subsets

The sub-set provisions may apply to sub-assemblies manufactured by one or more manufacturers. The essential requirements for cold meters shall also apply to the subsets of the cold meters. The following additional requirements apply:
7.1
Relative maximum tolerable error for flow sensor expressed as % in accuracy classes:
-
Class 2: E F = (2 + 0.02 q P /q), but not more than 5 per cent;
-
Class 3: E F = (3 + 0.05 q P /q), but not more than 5 %.
E F Is the difference between the specified value and the true value of the relationship between the output signal of the flow sensor and the mass or volume.
7.2
Relative maximum permissible error for pair of temperature sensors expressed as %:
-
E T = (0.5 + 3-∆ Q Min /∆ Q ).
E T Is the difference between the indicated value and the true value of the relationship between the output signal of the temperature sensor pair and the temperature difference.
7.3
Relative maximum tolerable error for calculation unit expressed as %:
-
E C = (0.5 + ∆ Q Min /∆ Q ).
E C Is the difference between the specified value of the thermal energy and the true value.
7.4
The limit value for a subset of a cold meter is equal to the absolute value of the maximum permissible error for the subset (see ch. 7.1, 7.2 or 7.3).
7.5
Subset subscriptions

Debit sensor:

-

Class of accuracy;

-

Flow limit values;

-

Temperature limit values;

-

Pulse value (p. Ex. Litre/pulse) or corresponding output signal;

-

Indication of flow direction;

Temperature Sensor Pair:

-

Sensor type (p. Ex. Pt 100);

-

Temperature limit values;

-

Temperature difference limit values;

Calculation unit:

-

Type of temperature sensors;

-

Temperature limit values;

-

Temperature difference limit values;

-

Pulse value (p. Ex. Litre/pulse) or corresponding input signal emitted by the flow sensor;

-

Flow sensor mounting mode: input or return.

8

Normative documents

The construction requirements and the metrological characteristics of cold meters shall be deemed to be observed if the meter meets the requirements contained in the following Swiss and European standard-setting documents:
-
SN EN 1434-1:2002, thermal energy meters-Part 1: general requirements;
-
SN EN 1434-4:2002, Thermal Energy Meters-Part 4: Examination for Model Approval.

State 1 Er January 2013