21/2001



DECREE



The Ministry of industry and trade



of 13 October. December 2000,



laying down the requirements for meters for liquids other than water

known by the EEC and on attachments to these meters



Change: 260/2003 Coll.



The Ministry of industry and trade shall determine in accordance with section 27 of Act No. 506/1990

Coll. on metrology, as amended by Act No. 121/2000 Coll. (hereinafter the "Act")

for the implementation of § 6 (1). 2 of the Act:



§ 1



This Decree lays down requirements for volumetric meters for liquids

other than water, in which the liquid causes the movement of mobile walls

measuring Chambers and which allow the measurement of any of the volumes. Volume

flow meters for liquids other than water, the device is made up of the measuring

device and the indicating device. The Decree also lays down requirements for the

ancillary equipment for meters for liquids other than water, for which the

shall be treated as zeroising mechanism for volume counters, totalizing counter

volume, volume, with more of the dial counter and multiple counters,

the price indicator and the printing device.



§ 2



Volumetric meters for liquids other than water can be instead of

official marks laid down special legal regulation ^ 1) marked

brand of the EEC pattern approval and initial verification mark, the

graphic design provides for a separate legal regulation, ^ 2) only if they meet the

the requirements set out in annex 1 to this notice, and which

It was verified the procedures laid down specific legislation. ^ 2)

Ancillary equipment for meters for liquids other than water can be

instead of official marks laid down special legal regulation ^ 1)

marked with the mark of EEC pattern approval and EEC initial verification, the

graphic design provides for a separate legal regulation, ^ 2) only if they meet the

the requirements set out in annex 2 to this Decree, and which

It was verified the procedures laid down specific legislation. ^ 2)



§ 3



This Decree shall take effect on the date of publication of the sectoral annex for

the measuring instrument to the Protocol to the Europe Agreement establishing an association between the

The Czech Republic on the one hand, and the European communities and their

Member States, of the other part, on conformity assessment and acceptance of

industrial products in the collection of international treaties, or on the date of entry

the Treaty of accession of the Czech Republic to the European Union entered into force,

If the day earlier.



Minister:



doc. Ing. Gregr, r.



Č. 1



Requirements for meters for liquids other than water



TECHNICAL SPECIFICATIONS FOR METERS FOR LIQUIDS OTHER THAN WATER



1.



The DEFINITION of the



1.1 the minimum delivery means the smallest volume of liquid, which can be

measured by a certain type of gauge.



1.2 the cyclic volume is the volume of liquid corresponding to the operating cycle

measuring equipment, IE. Summary of movements at the end of all

internal moving parts of the measuring device returns for the first time in their

initial positions.



1.3 Periodic variance means the maximum difference, during one

the working cycle between the volume displaced movable parts and

the corresponding volume shown on the indicating device that is

connected to a measuring instrument without play so that it points to the end of the cycle and for this

the cycle of the volume, which is equal to the cyclic volume. This difference can be reduced

a suitable calibration device.



1.4 the Adjustment is the activity of the measuring instrument to the listed

the State that is suitable for its use.



1.5 the adjustment device is part of the measuring device, which the device can be

appropriately set up to measure correctly.



2.



COUNTER



2.1 meters must incorporate a counter that shows the

the measured volume in cm3 or mm3, in l or dm3 or m3.



2.2 If the counter Has one or more elements, then the one whose scale is

marked with a minimum tick, is called the "first element".



2.3 conversion between the counter and the measuring device must be reliable,

durable and mediated by a mechanical connection or device with

permanent magnet.



2.4 Reading the data must be reliable, easy and unambiguous. If

the counter consists of several elements, as a whole, must be constructed in such a

to read data may be carried out by simple comparison of data from the

different elements.



2.5. The maximum capacity of the counter must match the value of 1 x 10 ^ n, 2 x

10 ^ n or 5 x 10 ^ n valid units of volume, n being a whole positive

or a negative number, or zero.



2.6 moving an element can either be a continuous (smooth) or discontinuous

(intermittent-increments).



2.7 if the movable part element has a continuous movement, the Division

the scale and the zero mark to allow the determination of the measured quantity for each

the position in which the element could stop.



2.8. The value scale intervals on the first element must match the value 1 x

10 ^ n, 2 x 10 ^ n or 5 x 10 ^ n valid units of volume.



2.9 the except element that corresponds to the maximum capacity of the counter, you must

one rotation of the element match the value 10 ^ n valid units of volume,

If it is fully visible graduated scale on this element.



2.10 if the element of a fixed circular scale and a rotating pointer,

must rotate indicator in a clockwise direction.



2.11 the counter with several elements must each turn moving parts

elements whose scale is fully visible, correspond to the value of the interval

the scale of the next element.



2.12 on the indicating device, with few elements must be an indication of the element with

a discontinuous movement, in addition to the first element to scroll by one figure,

While the preceeding element more than one-tenth of its

turns. This movement forward must cease when the previous element

shows zero.



2.13 it is to counter a few elements, and if the only see part of the

the scale of the second element and the elements following the movement of the other elements,

must be interrupted so that you can see in the window, the whole part of the relevant

element. Movement of the first element may be continuous or discontinuous.



2.14 if the indication given by the digits in a straight line and if the movement of the first

the element is discontinuous, it is permissible to indicate one or more fixed zero's

to the right of this element.



2.15 if in the window to see only part of the scale of the first element and

If this element has a continuous movement, may arise from the resulting

ambiguity in the reading, which should be deleted. For this reason, and in order to

to allow for reading by interpolation, may not have the appropriate box in parallel with

the movement of the scale dimension less than 1.5 times the distance between the

the Centre lines of two consecutive graduation marks, moving so

It is always seen at least two graduation marks, one of which will be

have a digit. Screen may be due to the zero mark asymmetric.



2.16 Marks on scales must have the same thickness, constant throughout the

the length of the tag, which does not exceed a quarter of the distance between the Centre

lines of two consecutive marks. Tags that correspond to the values

1 x 10 ^ n, 2 x 10 ^ n or 5 x 10 ^ n force units, are to be distinguished

the only difference in their lengths.



2.17 the distance between the axes of two consecutive graduation marks

shall not be less than 2 millimeters.



2.18 the height of the figures must not be less than 4 millimetres.



3.



ADJUSTMENT DEVICE



3.1 Meters must incorporate a calibrating device which can

alter the relationship between reported and actual volumes of liquids

passing through the meter.



3.2 If the adjustment device changes this relationship in a discontinuous manner, not after

the successive increments of this relationship never differ by more than 0.002

the values of the nominal flow.



3.3 Editing on a gauge enabling the bypass are disabled.



4.



SPECIAL CONDITIONS CONCERNING THE MINIMUM DELIVERY



4.1 the minimum delivery must be such that none of these values was not

greater than the maximum permissible error for such delivery, as

except as expressly stated in sections 10.2 and 10.3:



and the volume of the corresponding movement of two) millimetres on the scale of the first element

counter and one fifth of the value of the scale interval on the scale, if the first

the element has a continuous movement;



(b)) the volume corresponding to two consecutive movements of the digits, if

the first element has a discontinuous movement;



c) error that is normally caused by a clearance or offset

the drive between the measuring device and the first element of the counter;



d) twice the periodic variation.



4.2 when determining this minimum delivery should also be taken into account,

where necessary, the additional components on the measuring device,

in accordance with the requirements laid down in the Decree concerning this

the measuring device.



4.3 the minimum delivery must correspond to the values of 1 x 10 ^ n, 2 x 10 ^ n or 5 x

10 ^ n force units, n being a whole positive or negative number

or zero.



5.



THE MAXIMUM AND MINIMUM RATES OF FLOW



5.1 the maximum and minimum rates of flow are shown in the certificate of approval

a gauge type, in accordance with the results obtained when testing. A measuring instrument shall

be able to work after a period of time specified in the certificate of approval

a gauge type, of the EEC at a maximum flowrate without showed changes in the

its measurement characteristics.



5.2. the ratio between the maximum and minimum rates of flow must be equal to the value of the


10 for meters in General and to 5 for meters for liquefied gases.



6.



THE INFLUENCE OF PROPERTIES OF THE LIQUID, OF TEMPERATURE AND OF PRESSURE



6.1 the meter type approval certificate must specify the liquid or

liquids for which the meter is intended, the temperature limits for the liquid

to be measured, if these limits are below-10 ° C and above + 50

C, as well as the maximum working pressure.



6.2 Examination carried out for the purpose of type approval of the meter must show

the errors of the meter, caused by the maximum variation in the properties

fluids, pressure and temperature of the liquid, within the limits to be placed in the

the certificate of approval of the meter, do not exceed for each of these

half of the values of the factors set out in sections 10.1, 10.2 and 10.3.



7.



MAXIMUM PERMISSIBLE ERRORS ON A METER ALONE



7.1 if the initial verification of the gauge is preceded by the adjustment gauge

itself within the meaning of article 3, the maximum permissible errors at the time of this

the examination shall be equal to half of the maximum permissible errors laid down in

points 10.1, 10.2 and 10.3, but not less than 0.3% of the measured

the quantity is the same liquid for which the meter is intended.



7.2. However, if the measuring accuracy sufficient for the application of this

the request, the certificate of approval of a measuring instrument may increase

the maximum permissible errors, within the limits expressly referred to in points

10.1, 10.2 and 10.3.



7.3 the approval certificate of the gauge may reduce or alter the

the maximum permissible errors, in cases where the examination will be

perform only one of the liquids provided for or with a different liquid. In

the second case (that is, if the liquid used for authentication is different from

liquid measuring device designed for below) may, in the certificate

type approval of the meter indicate flow rates during testing, which differs from the

the flow of lying between the maximum and minimum flowrates.



8.



The INSCRIPTIONS



8.1 each meter must carry, in a legible and indelible manner, either on the

the dial counter, or on a special plate, the following information:



a) EEC type-approval mark,



(b)) the identification mark or name of the manufacturer,



(c)) any other manufacturer's designation,



d) serial number of meter and the year of manufacture,



(e)) the cyclic volume,



f) maximum and minimum flowrates,



g) maximum working pressure,



h) temperature range in the case that the liquid has to be measured at a temperature below

-10 ° C and above + 50 ° C,



I) nature of the liquid or liquids to be measured and the limits of viscosity,

kinematic or dynamic, where is not sufficient to determine their viscosity

an indication of the nature of the liquid.



8.2 on the dial counter must be clearly marked with the following information:



and) the unit in which the measured volumes are expressed or the symbol for this

units,



(b)) the minimum delivery.



8.3 where there is no ambiguity the direction of movement of the liquid must be

the direction indicated by the arrow on the housing of the meter.



5.2 on the gauges, which measure potable liquids and which can be

dismantled, the parts whose replacement can affect the results

the measurement must be repeated serial number on the last three digits of the

numbers.



8.5. The counter must have the type designation and the serial number.



9.



LOCATION OF SEALS AND VERIFICATION MARKS



9.1 the Seal must prevent access to parts that can alter

setting gauge, and must prevent the dismantling of the gauge, and even

partial, if not such dismantling in the certificate of approval

gauge type allowed (gauges, which measure potable liquids and which can be

disassembled).



9.2 on the measuring device, on the indicating device or on their coat

the critical section must be ensured a solid place that is visible to

without dismantling, for connection of the EEC verification marks.



9.3. the type-approval certificate of a measuring instrument may provide a place for the use of

the seals on the interchangeable parts of the gauge, which can be dismantled; This

the place is located right next to the serial number referred to in paragraph 5.2.



10.



MAXIMUM PERMISSIBLE ERRORS ON MEASURING SYSTEMS



10.1. If the meter Is incorporated in a measuring system, then when you first

Verify that the meter in normal conditions of use and to the extent

use, which are individually specified in the EEC certificate of approval

type, the maximum permissible errors specified in the following table as

negative or positive deviations depending on the measured quantity:

---------------------+--------------------------

Measured quantity | The maximum permissible errors

---------------------+--------------------------

From 0.02 to 0.1 liter | 2 ml

From 0.1 to 0.2 litre | 2% of the quantity measured

From 0.2 to 0.4 liter | 4 ml

From 0.4 to 1 litre | 1% of the quantity measured

From 1 to 2 litres | 10 ml

2 litres or more | 0.5% of the measured quantity

---------------------+--------------------------



10.2. The maximum permissible error of the maximum delivery is double the value of the

laid down in point 10.1 and independently of the measured quantity is never less

than permitted in the minimum delivery.



10.3 due to obzvláštních difficulties in the examination are the maximum

permissible error of measuring equipment for liquid gases or other

liquids measured at a temperature below-10 ° C or above + 50 ° C, and also

for devices for which there is no minimum flow rate greater than 1 litre per

an hour, equal to twice the errors referred to in points 10.1 and 10.2.



10.4 If there are any errors during the initial verification of the same sign,

at least one of them shall not exceed the limits laid down in paragraph 7.1.



Č. 2



THE REQUIREMENTS FOR ANCILLARY EQUIPMENT FOR METERS FOR LIQUIDS OTHER THAN

WATER



1.



THE ZEROISING MECHANISM FOR VOLUME COUNTER



1.1. The zeroising mechanism is the device by which the counter set to zero

either manually or automatically,



1.2. the zeroising mechanism must not affect the result of measurement.



1.3 If the reset operation may not be possible to indicate a new measured

the volume until the reset operation is complete.



1.4. the requirements referred to in points 1.2 and 1.3 are not mandatory



and for whose counter) bears the legend "direct public

the sale "or another equivalent restriction of use the same meaning;



(b)) for the counter for pointer-type mounted on meters whose maximum

rate of flow does not exceed 1200 litres per hour; If a measuring instrument intended to

public sale of liquids, may not be able to manually increase the display

volume.



1.5 to počítadlech a continuous movement, after each printing must not be

operation the permissible difference with respect to zero more than half of the maximum

permissible error on the minimum delivery marked on the indicator, but not

exceed one-fifth of the value of the scale interval. On the počítadlech with

a discontinuous movement, the information must be exactly zero.



2.



CUMULATIVE VOLUME COUNTER



2.1 counter with a zeroising mechanism may be associated with one or more

Summary počítadly, which show the sum of the different volumes at

recorded by itself.



2.2 total počítadlech must not be installed the zeroising mechanism.



2.3 the Totalizing counters can only be of the counter

the linear type.



2.4 Cumulative counter can be placed so that it is hidden.



2.5 unit in which the measured value of the cumulative counter (or its

tag) must be placed on the counter, and must comply with the requirements

special legislation. ^ 3)



2.6. the value of the scale of the first element of each cumulative counters

must match the value of 1 x 10 ^ n, 2 x 10 ^ n or 5 x 10 ^ n force

units of volume, n being a whole positive or negative number, or zero.

Must equal the value of the scale interval of the first element to include Ed with

a zeroising mechanism, or be greater than the scale interval.



2.7 If you can see both the summary and the information counter on the

counter with a zeroising mechanism, then the digits on the counter summary

they may not have dimensions greater than one half of the relevant dimensions of the

the counter with a zeroising mechanism.



3.



COUNTER WITH MORE VOLUME DIAL AND MULTIPLE COUNTERS



3.1 the counter may have several characteristics. In addition, it may

at the same time be accompanied by one or more concurrently showing the counters.



3.2 graduated in different počítadlech may be different,

but the minimum delivery must be uniform and must be determined as a function of

of the scale, as a result of the maximum values for this delivery.



3.3 Data on different ukazovatelích counter or counters shall not be

vary by more than the maximum permissible error on the minimum delivery

marked on the dial or ukazovatelích.



4.



THE PRICE INDICATOR



4.1 Linear displacement of the counter with a zeroising mechanism may be

supplemented by a straight-line computed-price linear with a zeroising mechanism,

the unit price is the price per unit of volume used to indicate

the measured volume.



4.2 the unit price must be adjustable. Elected as the unit price must

be displayed.



4.3. The mechanisms for selecting and displaying the unit price must be connected with the

the price indicator in such a way that the price corresponding to a measurement must always be equal

product of the selected and displayed unit price and the indicated volume.



4.4 requirements concerning volume counters must also apply by analogy to


calculated price counter, with the exception of item 1.5 relating to zero

device.



4.5 on the dial of the computed-price must appear used cash

unit or its symbol.



4.6 Size digits on the dial of the computed-price indicator must not exceed

measure the volume of the counter digits.



4.7. The zeroising mechanism to counter the price indicator and the volume must be

designed to work in a facility at one or the other of the two

the mechanism automatically return the second pointer to zero.



4.8



The computed price of a quantity equal to the maximum permissible error on the

minimum delivery marked on the dial of the counter must not be less

than one-fifth of the value of the scale interval, but not less than the price of

corresponding to an interval of 2 mm on the scale of the first element on the price indicator

If this element is continuous movement. This Division of one-fifth of the value or

two millimetres shall not, however, be less than the one of the

These monetary values, depending on the country where the

equipment used:



10 Belgian or Luxembourg centimes

,



1 French centime



1 Dutch cent



1 lira,



1 Pfennig,



10 cents.



The price of the quantity equal to the maximum permissible error on the minimum

delivery marked on the dial must be equal to at least two intervals on the

the scale, if the movement of the first price indicator is discontinuous. Graduation

However, the scale may not be less than one of the monetary values of the above

referred to.



4.9. The difference between the price indicated and the price calculated on the basis of

the unit price and the indicated quantity under normal working conditions

shall not exceed the price of the quantity equal to the maximum permissible error

When the minimum delivery marked on the dial of the indicating mechanism.



However, this difference need not be less than twice the one of

cash values listed in section 4.8.



4.10 To počítadlech a continuous movement, after each reset may not be

the tolerance from scratch more than half of the price for the quantity

equal to the maximum permissible error on the minimum delivery marked on the

the dial of the counters, but must not exceed one-fifth of the value of the interval

the scale.



However, this difference need not be less than one of the monetary values

listed in section 4.8.



On the počítadlech with a discontinuous movement, must be reset each time the

zero indication is displayed.



5.



THE PRINTING MECHANISM



5.1. to include Ed gauge can be connected digital printing

volume.



5.2. The value of the printed scale interval must match the 1 x 10 ^ n, 2 x

10 ^ n or 5 x 10 ^ n authorised volumetric units, n being a whole

a positive or negative number, or zero.



5.3. the value of the printed scale interval must not be greater than the

maximum permissible error on the minimum delivery marked on the dial

the indicating device.



5.4 the value of the printed scale interval must be marked on the printing

device.



5.5 printed quantity must be expressed in one of the

volumetric units.



The printing mechanism must, on a ticket printed digits including decimal

commas used unit or its brand.



5.6 the printing mechanism may print information regarding the benefits, such as

for example, serial number, date, place of measurement, nature of the liquid.



5.7 Printing device may be so designed that the printing may

repeated. In this case, the printed records must fully agree and have

the same serial number.



5.8 if the volume determined by the difference between two printed values,

one of them can be expressed as zero, it shall not be possible to remove the

the ticket from the juxtaposition of the device during the measurement.



5.9. The printing mechanism, except in the case referred to in point 3.6, you must

have a zeroising mechanism combined with that device counter.



5.10. The difference between the quantity indicated and the quantity printed must not

exceed the value of one printed scale interval.



5.11 the printing mechanism may print in addition to the measured volume of either the price per

This volume, or this price and the unit price. If the printing mechanism

associated with the volume and the price indicator for direct public

the sale, then you may also print only the price to be paid.



Digits including decimal point, used the cash unit or its

the symbol must be printed on the ticket. The printed digits for price

they may not have dimensions greater than the dimensions of the printed digits for

the measured volume.



5.12. The value of the printed scale interval must match the value for the price

1 x 10 ^ n, 2 x 10 ^ n or 5 x 10 ^ n monetary units, n being a whole

a positive or negative number, or zero.



This value shall not exceed the price of the quantity equal to the maximum

permissible error on the minimum delivery marked on the dial

the indicating device.



The value of the scale interval, however, may not be less than one of the

cash values listed in section 4.8.



5.13. If the meter is fitted with a price indicator the difference between displayed

the price and the printed price must not exceed the value of the printed scale interval

the scale.



If the meter is fitted with a price indicator, the difference between the printed

price and the price calculated from the indicated quantity and the unit price

must satisfy the conditions laid down in point 4.9.



6.



PRE-SETTING MECHANISMS



6.1 the meter may be equipped with a device for the preset.



Pre-setting mechanisms are devices that allow you to select quantity

the measure, which automatically stops the flow of liquid, if it was measured

the selected quantity.



6.2 the selected quantity is displayed by using a device with a scale and zero

line or on a numerical device.



6.3. Where pre-setting can be done by using several independent controls,

the value of the scale interval for one control must equal the scope of the

driver preferences for the next lower decade.



6.4. Pre-setting mechanisms may be so arranged that repetition of the

the selected quantity does not require a new setting of the controls.



6.5 where the displayed figures of the pre-setting mechanisms are separate from the

counter digits and can be seen at the same time, numbers, the equipment for

preference shall not be more than three-quarters of the corresponding

dimensions of the counter.



6.6 the information of the selected quantity may, during measurement, either remain unchanged or

gradually return to zero.



4.2 the difference found under normal working conditions between the selected

volume and quantity indicated at the end of the measuring operation must not exceed

half of the maximum permissible error on the minimum delivery.



4.2 the selected quantities and the volume display counter must be expressed in the same

unit. This unit (or its symbol) must be on the device for

the preset volume marked.



4.3 value of the smallest scale interval of the pre-setting mechanisms

shall not be less than the value of the scale interval on the first element of the counter.



6.10. Pre-setting mechanisms may incorporate a mechanism for

fast stopping of the flow of the liquid, if necessary.



6.11 where a pre-setting mechanism built into the device for

the regulation of the flow restriction at the end of the measurement, the device must be provided

the seal is damaged, if it is necessary, in order to prevent any change to the selected

settings.



6.12. the requirements of items 6.7 and 6.11 do not apply, if it is associated with the measuring instrument

the printing mechanism, which allows you to issue a printed ticket or if the

the case of direct public sale device hidden.



6.13 the gauge to the price indicator can also be fitted for

preset prices. In this case, the flow of the fluid stops at that moment,

When the issued volume of liquid corresponding to the chosen price.



7.



SEALING



In order to avoid a withdrawal of the ancillary equipment and access to parts

the gauge, which can be set, the gauge must be fitted with plugs.



1) Decree No. 262/2000 Coll., which ensures consistency and accuracy

measuring instruments and measurements.



2) Decree No. 332/2000 Coll., laying down some of the procedures in the

type-approval and verification of measuring instruments, called marker set

EEC.



3) Decree No 264/2000 Coll., on the basic units of measurement and

other units and their labelling.