902 KAR 100:145.
Cabinet systems.
RELATES TO: KRS
211.842-211.852, 211.990(4)
STATUTORY
AUTHORITY: KRS 194.050, 211.090, 211.844
NECESSITY,
FUNCTION, AND CONFORMITY: The Cabinet for Human Resources is authorized by KRS
211.844 to regulate the possession or use of sources of ionizing or electronic
product radiation and to regulate the handling and disposal of radioactive
waste. The purpose of this administrative regulation is to provide requirements
for the possession, use or operation of cabinet x-ray systems.
Section 1.
Applicability. This administrative regulation shall apply to persons who
possess, use or operate cabinet x-ray systems. This administrative regulation
does not apply to microscopic analytical x-ray systems.
Section 2. For
cabinet x-ray systems installed prior to April 10, 1975, the systems shall:
(1) Have the
radiation machine and objects exposed thereto, within a permanent enclosure,
within which no person is permitted to remain during the generation of
x-radiation.
(2) Have adequate
interlocks provided so that if doors or panels to the enclosure are opened, the
radiation-producing machine is shut off automatically. After shutoff, it shall
be possible to restore the machine to full operation only from the control
panel outside the room.
(3) If the
enclosure is of a size or is so arranged that the operator cannot readily
determine whether the enclosure is unoccupied, there shall be provided:
(a) Audible or
visible warning signals within the enclosure which are required to be activated
before irradiation can be started.
(b) Suitable
means of exit, so that a person who accidentally may be shut in can leave the
enclosure without delay, or effective means within the enclosure for preventing
or quickly interrupting the irradiation, and which cannot be reset from outside
the enclosure.
(4) Except as
provided in subsection (5) of this section, the exposure at accessible regions
two (2) inches from the outside surface of the enclosure cannot exceed
five-tenths (0.5) milliroentgen in one (1) hour.
(5) If the
registrant has submitted operating and safety procedures, and the cabinet has
approved the procedures, which can show that an individual cannot receive a
radiation dose in excess of the limits specified in these administrative
regulations, the radiation levels specified in subsection (4) of this section
may be exceeded. However, cabinet x-ray systems under this subsection shall not
exceed the limits specified below:
(a) The exposure
at accessible and occupied areas one (1) foot from the outside surface of the
enclosure shall not exceed ten (10) milliroentgens in one (1) hour.
(b) The exposure
at accessible and normally unoccupied areas one (1) foot from the outside
surface of the enclosure shall not exceed 100 milliroentgens in one (1) hour.
(6) Operating
procedures.
(a) Before a new
installation is placed in routine operation, a radiation protection survey
shall be made. This shall also be done after changes in an existing
installation which might affect its radiation safety.
(b) A copy of
each radiation protection survey shall be signed and dated by the surveyor, and
kept on file by the individual in charge of the installation.
(c) The
installation shall be operated in conformance with recommendations of the
protection survey.
(d) The
registrant shall designate a competent employee as the radiation safety
officer.
Section 3. For
cabinet x-ray systems installed on or after April 10, 1975, systems shall:
(1) Have the
radiation machine and objects exposed thereto, within a permanent enclosure,
within which no person is permitted to remain during the generation of
x-radiation.
(2) Be so
constructed that radiation emitted from the cabinet x-ray system shall not
exceed an exposure of five-tenths (0.5) milliroentgen in one (1) hour at points
five (5) centimeters outside the external surface.
(3) Have a
permanent floor. Support surfaces to which a cabinet x-ray system is
permanently affixed may be deemed the floor of the system.
(4) Be so
constructed that the insertion of part of the human body through ports or
apertures shall not be possible.
(5) Be so
constructed that each door of a cabinet x-ray system has a minimum of two (2)
safety interlocks. One (1), but not both of the required interlocks, shall be
configured in a manner that door opening results in physical disconnection of
the energy supply circuit to the high-voltage generator, and this disconnection
shall not be dependent upon moving parts other than the door.
(6) Be so
constructed that each access panel shall have at least one (1) safety
interlock.
(7) Following
interruption of x-ray generation by the functioning of a safety interlock, use
of a control provided in accordance with subsection (10)(b) of this section
shall be necessary for resumption of x-ray generation.
(8) Be so
constructed that failure of a single component of the cabinet x-ray system
shall not cause failure of more than one (1) required safety interlock.
(9) Be so
constructed that a ground fault shall not result in the generation of x-rays.
(10) For systems
to which this section is applicable, there shall be provided:
(a) A
key-actuated control to insure that x-ray generation is not possible with the
key removed.
(b) A control or
controls to initiate and terminate the generation of x-rays other than by functioning
of a safety interlock of the main power control.
(c) Two (2)
independent means which indicate if and only if x-rays are being generated,
unless the x-ray generation period is less than one-half (1/2) second, then the
indicators shall be activated for one-half (1/2) second, and which are
discernible from points at which initiation of x-ray generation is possible.
Failure of a single component of the cabinet x-ray system shall not cause
failure of both indicators to perform their intended function. One (1), but not
both, of the indicators required by this paragraph may be a milliammeter
labeled to indicate x-ray tube current. Other indicators shall be legibly
labeled "X-RAY ON."
(d) Additional
means other than milliammeters which indicate if and only if x-rays are being
generated, unless the x-ray generation period is less than one-half (1/2)
second then the indicators shall be activated for one-half (1/2) second, to
insure that at least one (1) indicator is visible from each door, access panel,
and port, and is legibly labeled "X-RAY ON."
(11) For cabinet
x-ray systems designed to admit humans, there shall also be provided:
(a) A control
within the cabinet for preventing and terminating x-ray generation, which
cannot be reset, overridden or bypassed from the outside of the cabinet.
(b) No means by
which x-ray generation can be initiated from within the cabinet.
(c) Audible and
visible warning signals within the cabinet which are activated for at least ten
(10) seconds immediately prior to the first initiation of x-ray generation
after closing doors designed to admit humans. Failure of a single component of
the cabinet x-ray system shall not cause failure of both the audible and
visible warning signals.
(d) A visible
warning signal within the cabinet which remains actuated if and only if x-rays
are being generated, unless the x-ray period is less than one-half (1/2) second
then the indicators shall be activated for one-half (1/2) second.
(e) Signs
indicating the meaning of the warning signals specified in subsection (11)(c)
and (d) of this section and containing instructions for the use of the control
specified in subsection (11)(a) of this section. These signs shall be legible,
accessible to view, and illuminated if the main power control is in the
"on" position.
(f) The
designated operator(s) shall have a copy of the manufacturer's operating
procedures and maintenance manual for the equipment being used and shall have
been instructed in the proper techniques of equipment utilization.
(g) Tests for
proper operation of interlock and door seal systems shall be conducted and
recorded at intervals not to exceed one (1) month.
(12) Have
permanently affixed or inscribed on the cabinet x-ray system at the location of
controls which can be used to initiate x-ray generation, a clearly legible and
visible label bearing the statement: "CAUTION: X-RAYS PRODUCED WHEN ENERGIZED."
(13) Have
permanently affixed or inscribed on the cabinet x-ray system adjacent to each
port a clearly legible and visible label bearing the statement: "CAUTION:
DO NOT INSERT ANY PART OF THE BODY WHEN SYSTEM IS ENERGIZED - X-RAY HAZARD."
(14) Additional
requirements for x-ray baggage inspection systems. X-ray systems designed
primarily for the inspection of carryon baggage at airline, railroad, and bus
terminals, and at similar facilities, shall be provided with means specified in
paragraphs (a) and (b) of this subsection, to insure operator presence at the
control area in a position which permits surveillance of the ports and doors
during generation of x-radiation.
(a) During an
exposure or preset succession of exposures of one-half (1/2) second or greater
duration, the means provided shall enable the operator to terminate the
exposure or preset succession of exposures at once.
(b) During an
exposure or preset succession of exposures of less than one-half (1/2) second
duration, the means provided may allow completion of the exposure in progress
but shall enable the operator to prevent additional exposures. (1 Ky.R. 417;
eff. 2-5-75; Am. 12 Ky.R. 1412; eff. 3-4-86; 18 Ky.R. 1570; eff. 1-10-92.)